Insufflator testing.

Insufflator testing.

With the insufflation tubing kinked, the intraabdominal pressure should rapidly rise (e.g., 30mmHg), thereby exceeding the
preset 15mmHg pressure set point. The flow of CO2 should immediately cease
(0 L/min) and an alarm should sound.